Liquid NanoParticle Sizer: Safety and Handling Procedures
Liquid NanoParticle Sizer User Guide
3
Version 1.0.1
Warnings
Please familiarize yourself with the following warnings before operating the
Liquid NanoParticle Sizer:
▪
The Liquid NanoParticle Sizer
must be used following manufacturer’s
specifications otherwise safety cannot be guaranteed.
▪
All service work must be performed by qualified service technicians -
only qualified service technicians should remove the cover.
▪
When the nebulizer is running, there are hot surfaces inside the
device. Do not remove the cover at any time unless you are a qualified
service technician.
▪
To prevent electric shocks, ensure that all electrical outlets are
grounded.
▪
The aerosol particles created by the Model 9310 Liquid NanoParticle
Sizer may pose a health risk if inhaled. If not connected to other
instrumentation, vent the aerosol output to a fume hood.
▪
Follow the instructions for all inlet and outlet connections. Incorrect
connections will cause the nebulizer to malfunction.
▪
The air or nitrogen supplied to the nebulizer must be clean, dried, oil-
free and regulated at 50-60 psi.
▪
During normal operation, do not tilt the nebulizer at angles >10
.
▪
You must drain the nebulizer before you move or ship it. Do not ship
an undried/undrained nebulizer back to Kanomax FMT, Inc.: doing so
might damage the device and invalidate the warranty.
▪
Do not subject an undrained nebulizer to freezing temperatures:
doing so might damage the device and invalidate the warranty.