4.3.15.1 Modes to be tested
Note:
When setting the setpoints, always select the maximum effect.
4.4 Testing electrical safety
Note:
Test electrical safety with the line voltage and frequency stated on the manufacturer’s
identification plate.
Warning: Do not touch the device during testing!
Note:
Make sure that the relay contacts for the output sockets are closed for testing the leakage
currents.
To close the relay contacts, proceed as follows:
– Connect the power cord to the device.
– Turn the device on.
– Press the arrow button in the upper right-hand corner of the Procedure Overview screen
.
– Press the “Service Programs: Select Option” button in the “Service Programs” menu.
– Enter the service password. The password when the device leaves the manufacturers is “Tuttlingen”.
– Use the acknowledgement button in the upper right-hand corner of the screen to confirm the
password.
– Press the “Output relays on / off” button.
The closed relays are acknowledged by a green dot on the button.
Note:
When a test device is used for electrical safety, power is supplied to the device from the test
device. When selecting the program for the measurements you should make sure that the test
device enters a wait state after it has switched on the HF generator in order that the steps
listed above can be carried out.
The test device may feature an automatic or semi-automatic mode which only starts after
additional acknowledgement by the test program.
4-12
12.06 / V1.0
AUTOCON
®
II 400
MODEL 205352 20-1
Mode
Power
Effect
Monopolar Cut
Power Cut
300 W ±20%
Effect 8 on RL 500
Top Cut
300 W ±20%
Effect 8 on RL 500
C-Cut
200 W ±20%
Effect 4 on RL 500
Bipolar Cut
Bipolar Cut
100 W ±20%
Effect 8 on RL 500
Saline C-Cut
370 W +9% / -20%
Effect 8 on RL 500
Monopolar Coagulation
Standard Coag
200 W ±20%
Effect 8 on RL 50
Forced Coag
120 W ±20%
Effect 4 on RL 500
Spray Coag
120 W ±20%
Effect 2 on RL 500
Bipolar Coagulation
Bipolar Soft Coag
120 W ±20%
Effect 8 on RL 75
Saline Coag
200 W ±20%
Effect 8 on RL 500
Summary of Contents for 205352 20-1
Page 8: ......
Page 9: ...AUTOCON II 400 AUTOCON II 400 AUTOCON II 400 20 5352 20 11x 20 5352 20 11x 20 5352 20 11x ...
Page 10: ......
Page 13: ...III Geräteabbildungen Photographs of the equipment Imágenes del equipo ...
Page 14: ...IV W R Q Geräteabbildungen Photographs of the equipment Imágenes del equipo I O Y U T E ...
Page 146: ......
Page 152: ......
Page 156: ......
Page 158: ......
Page 172: ......
Page 186: ...Figure 4 11 Figure 4 12 Figure 4 13 4 14 12 06 V1 0 AUTOCON II 400 MODEL 205352 20 1 ...
Page 192: ......
Page 198: ......
Page 200: ......
Page 202: ......
Page 203: ......