12.06 / V1.0
AUTOCON
®
II 400
MODEL 205352 20-1
Article number
Device type
Serial number
Software versions
Generator control
ESI
User interface
SCB
Visual inspections
Labeling
Potential equalization
Fuse type (T 8.0 AH / T 4.0 AH)
Adhesive label
Fuse labels
Serial number
Manufacturer’s identification plate
CE mark
TÜV / NRTL
Components
Feedthrough cover
Power switch
green when on
Socket modules
undamaged
Frame
undamaged
Housing
undamaged
Electrical safety
Protective ground
resistance (with
power cord)
0.20
Earth leakage
current
N.C.:
0.50 mA
S.F.C.:
1.00 mA
Housing leakage
current
N.C.:
0.10 mA
S.F.C.:
0.50 mA
Patient leakage
current
N.C.:
0.01 mA
S.F.C.:
0.05 mA
Patient leakage
current (with line
voltage on applied
part)
S.F.C.:
0.05 mA
Protective conductor testing
Protective conductor contact, power
socket against housing
Housing base
Cover
Front panel
Front frame
PE contact feedthrough (3-pole)
Bus termination testing SCB
2)
Socket 1
one plug connected
Socket 1
both plugs
connected
Socket 2
one plug connected
Socket 2
both plugs
connected
Device startup
Procedure list
Alive signal (clock)
Footswitch
recognition
Message detected
Message separated
Activation signal
Bipolar Cut
Bipolar Coag
Procedure switchover A / B
3)
Only triple-pedal footswitch
Handpiece detection
4)
Activation signal
Cut
Coag
Device startup – Error detection
Touch screen
pressed
Error message
Footswitch
pressed
Error message
Handpiece
pressed
Error message
NE monitoring
Limit resistance
single-faced
Limit resistance
double-faced
Symmetry
double-faced
Autostart bipolar
Autostop bipolar
Instrument detection MF socket
Power measurement cut monopolar
Power Cut
300 W ±20%
Top Cut
300 W ±20%
C-Cut
200 W ±20%
Power measurement cut bipolar
Bipolar Cut
100 W ±20%
Saline C-Cut (MF)
370 W +9% / -20%
Power measurement, monopolar
coagulation
Standard Coag
200 W ±20%
Forced Coag
120 W ±20%
Spray Coag
120 W ±20%
Power measurement, bipolar
coagulation
Bipolar Soft Coag
120 W ±20%
Saline Coag (MF)
Date for next safety check
Test report – safety checks
2) Tested only when the device is linked up to KARL STORZ-SCB
®
.
3) Tested only when a triple-pedal footswitch is installed.
4) When a handpiece 205300 43 and high-frequency cable 205300 45 are used.
Summary of Contents for 205352 20-1
Page 8: ......
Page 9: ...AUTOCON II 400 AUTOCON II 400 AUTOCON II 400 20 5352 20 11x 20 5352 20 11x 20 5352 20 11x ...
Page 10: ......
Page 13: ...III Geräteabbildungen Photographs of the equipment Imágenes del equipo ...
Page 14: ...IV W R Q Geräteabbildungen Photographs of the equipment Imágenes del equipo I O Y U T E ...
Page 146: ......
Page 152: ......
Page 156: ......
Page 158: ......
Page 172: ......
Page 186: ...Figure 4 11 Figure 4 12 Figure 4 13 4 14 12 06 V1 0 AUTOCON II 400 MODEL 205352 20 1 ...
Page 192: ......
Page 198: ......
Page 200: ......
Page 202: ......
Page 203: ......