154
Anhang
Hinweise zur elektromagnetischen
Verträglichkeit (EMV)
Anexo
Indicaciones sobre compatibilidad
electromagnética (CEM)
Appendix
Electromagnetic Compatibility (EMC)
Information
Table 204
Guidance and manufacturer’s declaration – electromagnetic immunity –
for equipment and systems that are not life-supporting
IMAGE 1 HUB™ HD (22201020-1xx) and IMAGE 1 Pure HD (22202020-yyy) are intended for use in the electromagnetic environment speci
fi
ed below.
The customer or user of the IMAGE 1 HUB™ HD (22201020-1xx) and IMAGE 1 Pure HD (22202020-yyy) should ensure that they are used in such an
environment.
Immunity test
EN/IEC 60601
test level
Compliance level
Electromagnetic environment - guidance
Portable and mobile RF communications equipment should be used
no closer to any part of the IMAGE 1 HUB™ HD/IMAGE 1 Pure HD,
includ ing cables, than the recommended separation distance calcu-
lated from the equation applicable to the frequency of the transmitter.
Recommended separation distance:
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to < 80 MHz
3
Vrms
150 kHz to < 80 MHz
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3
V/m
80 MHz to < 800 MHz
800 MHz to 2.5 GHz
Where P is the maximum output power rating of the transmitter in
watts [W] according to the transmitter manufacturer and d is the
recommended separation in meters [m].
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey
a
, should be less than the compliance
level in each frequency range
b
.
Interference may occur in the vicinity of equipment marked with
the following symbol:
Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and re
fl
ection by structures, objects
and people.
a
Field strengths from fixed transmitters, such as base stations for radio [cellular/cordless] telephones and land mobile radios, amateur radio, AM and
FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the unit is used exceeds the
applicable RF compliance level above, the unit should be observed to verify normal operation. If abnormal performance is observed, additional mea-
sures may be necessary, such as reorienting or relocating the unit.
b
Over the frequency range 150 kHz to 80 MHz,
fi
eld strengths should be less than
3
V/m.
[
]
P
d
3
/
5
.
3
=
[
]
P
d
3
/
5
.
3
=
[ ]
P
d
3
/
7
=
Summary of Contents for Image 1 HUB
Page 2: ......
Page 5: ...III Geräteabbildungen Photographs of the equipment Imágenes del equipo ...
Page 170: ......
Page 174: ......
Page 175: ......