Technische
Beschreibung
Technical
description
Техническое
описание
Error message
Cause
Remedy
1 = E01: MEMORY ERROR
Parameters could not be saved successfully
Turn device on/off.
Load default values.
Or send unit in for repair
2 = E02: MOTORBOARD DEFECTIVE
Motor control overloaded
Turn device on/off.
Or send unit in for repair
3 = E03: HARDWARE CONFIGURATION ERROR
Hardware configuration is incorrect
Send unit in for repair.
4 = E04: SOFTWARE CONFIGURATION ERROR
Software configuration is incorrect
Send unit in for repair.
5 = E05: –
6 = E06: –
7 = E07: –
8 = E08: –
9 = E09: MOTOR DEFECTIVE M
Motor on channel 1 is defective
Release footswitch and restart.
If the error persists, a different handpiece
must be used.
10 = E10: -
11 = E11: NO FOOTSWITCH
Footswitch not connected correctly
Connect footswitch.
If the error persists, check whether the
error is on the footswitch or unit.
Send the defective component in for repair.
12 = E12: NO MOTOR
No motor/handpiece connected
Connect motor/handpiece.
13 = E13: MOTOR OVERHEATING M
Motor/handpiece on channel 1 overheated
Allow the connected handpiece to cool
down or operate an additional handpiece
on the 2nd channel.
14 = E14: -
15 = E15: -
16 = E16: -
17 = E17: MOTOR NOT SUPPORTED M
The motor/handpiece connected on channel 1 is not
supported.
Only use handpieces approved by KARL
STORZ. A service update may be neces-
sary.
18 = E18: -
19 = E19: MOTOR STOPPED M
Motor/handpiece on channel 1 blocked in operation
Release footswitch and restart.
If the error persists, clean the handpiece or
exchange the attachment.
20 = E20: -
9. 2 UNIDRIVE
®
troubleshooting
table
1
NOTE:
If the device is not working as requi-
red, check the error message displayed in
the 7-segment display. If necessary, contact
the KARL STORZ Service.
50