22.2. VR2 Controller, Battery & Diagnostics Indicator
The following table indicates what the gauge will display for any given state.
Indication
Status
Meaning
What to Do
1 signal
The battery needs charging
or there is a bad connection
to the battery.
Check the connections to the battery. If
the connections are good, try charging
the battery.
2 signals
The left hand motor has a
bad connection.
Check the connections to the left hand
motor.
3 signals
The left hand motor has a
short circuit to a battery
connection.
Connect your service agent.
4 signals
The right hand motor has a
bad connection.
Check the connections to the left hand
motor.
5 signals
The left hand motor has a
short circuit to a battery
connection.
Connect your service agent.
6 signals
The wheelchair is being
prevented from driving by
an external signal.
Please remove the charger.
7 signals
A joystick fault is indicated. Make sure that the joystick is in the
center position.
8 signals
A joystick fault is indicated. Make sure that all connections are
secure.
9 signals
The parking brakes have a
bad connection.
Check the parking brake and motor
connections and make sure all
connections are secure.
10 signals
An excessive voltage has
been applied to the control
system.
Check the battery connections.
7 signals
+ S
A communication fault is
indicated.
Make sure that joystick cable is
securely connected and not damaged
8 signals
+ A
An actuator trip is indicated. If more than one actuator is fitted
check which actuator is not working
correctly. Check the actuator wiring.