45
6.
Now remove the top ring, and the five ceramic top plates (see picture below).
7.
Remove the inner “Black Ceramic” cylinder that protects the element.
8.
Gently pull the element upwards.
Heating element installation
From the instructions above, the crucible heater is in a partially disassembled
state. You are ready to install the new heating element.
1.
Make sure the instrument is still disconnected from its power outlet.
2.
Wear surgical-style gloves. They will protect you from the insulation fibers,
and will protect the heating element’s clear tubes from your skin oil.
Heating
element
Black ceramic
cylinder
Ceramic plates
Tip
Summary of Contents for K1 PRIME