52
4.
Touch “Download”.
5.
A dialog will appear, requesting a password and what data to download. The
password is “111111” (six times the digit one). You may have to drag the
dialog window to the left, to show the keyboard and then type the password.
6.
Select “Download Project Files” if you want to restore or upgrade the
firmware (i.e. the instrument’s operating system).
7.
Select “Download History Files” if you want to restore the fusion programs’
parameters. Note that this will erase ALL fusion programs currently stored on
your fluxer.
8.
Touch “OK”.
9.
Now, choose where you want the data to be fetched. By default, you will
have to double-click on “USBDISK”, then click on its sub-directory, named
“disk_a_1”. If the data was created in another (sub-)directory, you will need
to browse to reach it. In particular, firmwares obtained from Katanax (via
Web or email) are located in a sub-directory named “Firmware K1P…” ; you
must select (highlight) this folder at this step.
10.
Click “OK” to start the transfer to the USB drive. The screen will black out,
and the fluxer will re-boot.
11.
The new firmware or programs are loaded, and you can remove the USB
drive.