4
Installation
Congratulations on your acquisition of the K1 Prime fluxer, from Katanax. Please
read the following section for proper commissioning of your instrument. Do not
hesitate to contact Katanax with any question you might have with this crucial
step.
IMPORTANT
: It is advisable that at least two persons carry this instrument to
avoid injuries. Do not drop instrument.
Box contents
The instrument comes with its essential accessories. In addition to optional items
you might have ordered, the box should contain:
1 fluxer (K1 Prime main unit)
1 mold heating/cooling unit
1 mold holder to fit your mold size
1 power cord
1 tool kit (hex keys, spare fuses...)
1 instruction manual (this booklet)
Additionally, if you have ordered a K1 Prime with solution-making capability, you
will find:
1 solution agitation module
1 unbreakable Teflon beaker
1 magnetic stirring bars
Location
Vent hood
Molten flux, additives and some samples may produce vapors and gases that
need to be extracted, but since this instrument does not produce post-combustion
hot gases, only a small hood is required. A wide, funnel-ended, flexible suction
duct may be the most practical and economical solution.
Table
The counter on which the instrument is to be used must be made of non-
combustible material and must be able to withstand a weight of 30 kg (66 lbs). It
is recommended to use the instrument from a standing position, and adjust table
height accordingly.
It is advisable that at two persons carry the instrument to avoid back injuries.
Clearances
Because your new fluxer will produce heat, a minimal safety clearance must be
provided to prevent surrounding material from heating and potentially catching
fire. All around the instrument, a minimum distance of 30 cm (12’’) must be free
Summary of Contents for K1 PRIME