7.2.4 Cleaning: Automated internal cleaning
KaVo recommends thermodisinfectors in accordance with EN ISO 15883
that are operated with alkaline cleaning agents at a pH of max. 10 (e.g.
Miele G 7781 / G 7881 – validation was performed with the ""VARIO-TD"
programme, "neodisher
®
mediclean" cleaning agent, "neodisher
®
Z" neu‐
tralisation agent, and "neodisher
®
mielclear" rinsing agent and extends only
to the compatibility of materials with respect to KaVo products).
▶
For program settings as well as cleansers and disinfectants to be used,
please refer to the Instructions for Use of the thermodisinfector (com‐
plying with max. pH value of 10).
▶
In order to prevent negative effects on the medical device, make sure
that the interior and the exterior of the medical device are dry, and then
lubricate immediately with care agents from the KaVo care system.
Setup methods according to DIN EN ISO 17664
40