To prevent collisions during movements of the patient simulator, a safety shutoff ① is
installed on the upper front of the unit to protect people from injury. When the safety
shutoff ① is activated, the patient simulator immediately stops moving.
An exclamation mark is shown on the display.
S1
S2
S3
S4
▶ Press the S2 button to display the message.
▶ To deactivate an activated safety shutoff, remove the triggers from the motion
range of the patient simulator and re-set the end-positions of the patient simulator.
+
▶ Press and hold down the automatic position 1 and automatic position 2 buttons
simultaneously and also press the park position button briefly.