6 Service and Care
6.1 Servicing
DANGER
Electrical power.
Electrical shock
▶ If the light is attached, pull the mains plug of the unit and of the light or disconnect
all poles of the unit from the voltage source.
Note
Repair and servicing work on the electrical part of the unit may be done only by
skilled staff or by staff trained in the factory.
During servicing work, pull the mains plug out of the socket or disconnect the unit
from the mains.
6.1.1 Draining and cleaning the wastewater tank (suction)
Note
The wastewater tank ① contains a float that closes the connection to the suction
system when the liquid level is high.
Before use, check the filling level of the wastewater container and empty it, if
needed.
▶ Remove the wastewater tank ① and properly dispose of the content.
① Wastewater tank
② Pressure tank with holding magnets
▶ Rinse the wastewater tank ① thoroughly.
▶ Insert the waste water tank ① along the left inside of the aluminium profile until
the wastewater tank touches against the housing in the back.
Note
Daily drain and clean the water tank in the suction unit.