8 Troubleshooting
Malfunction
Cause
Remedy
No cold light on the hand-
pieces.
Cold light not preselected.
▶ Preselect cold light.
The high-pressure lamp or
Multi LED on the hand-
piece is defective.
▶ Replace the high-pressure lamp or Multi LED. also
refer to: Instructions for Use of the handpiece
No spray in the hand-
pieces.
No spray preselected.
▶ Preselect spray.
Check setting on butterfly valves of the dentist ele-
ment.
Close the ring for con-
trolling the spray on the
handpieces.
▶ Open the ring for controlling the spray on the hand-
pieces.
Pressure tank is not inser-
ted correctly.
▶ Insert the pressure tank correctly until the clicking
of the holder magnets can be heard.
69 / 74