3 Product description | 3.5 Technical Specifications
3.5 Technical Specifications
Space needs of DSEclinical type 5198
Width
approximately 486 mm
Depth
approximately 250 mm
Maximum height
2300 mm
Weight
Approximately 70 kg
Rated voltage
Nominal voltage
(country specific)
100 V / 50/60 Hz
120 V / 50/60 Hz
230 V / 50/60 Hz
Mains fuse
Type
T6.3 H / 250 V
Maximum permissible mains voltage
fluctuations
±10 % of the nominal voltage
Overvoltage category
II
Customer-provided fuse protection
Automat C16 or screw-plug fuse 10
A
Protective conductor above floor
See DIN VDE 0100-710, 1000 mm
Triple function handpiece One
Flush the water and air passages for 20 to 30 seconds before working at the
beginning of the day.
Water pressure
1.5 ± 0.3 bar; flow pressure; probe
tip (1.009.6634)
Water flow
90 ± 15 ml/min
Air pressure
1.5 ± 0.2 bar; flow pressure; probe
tip (1.009.6634)
Air flow
15 ± 3 Nl/min
Triple function handpiece
Flush the water and air passages for 20 to 30 seconds before working at the
beginning of the day.
23 / 100