Instructions for use DSEclinical Typ 5198
3 Product description | 3.5 Technical Specifications
24 / 100
Water pressure
1.5 ± 0.3 bar; Flow pressure; 4 x
manometer
Max. static pressure water
2.5 ± 0.3 bar
Water flow
80 ± 10 ml/min
Air pressure
3.3 ± 0.1 bar; Flow pressure; 4 x
manometer
Max. static pressure air
4 + 0.5 bar
Air flow
at least 16 Nl/min
Media
Water quality
Tap water
Water hardness
1.5 to 2.14 mmol/l
≙
8.4 to 12° dH
pH
7.2 to 7.8
Connected loads
Power consumption
800 VA
Water pressure at customer
premises
2 to 6 bar
Air inlet pressure
5.2 bar to 7 bar
Maximum water consumption
0.1 l/min
Maximum air consumption with ex-
ternal suction
60 l/min
Mode
Intermittent operation for Motor K
200 / Motor KL 703
Operating time is 2 minute; pause is
8 minutes
Intermittent operation of KL 702 mo-
tor
Operating time is 0.5 minute; pause
is 9 minutes
Intermittent operation of lift motor
simulator
Operating time is 0.5 minute; pause
is 7 minutes
Speed
Dental drives
Instruments depending on type of
unit
up to 300,000 rpm
KL motor
40,000 rpm