4 Commissioning
4 Commissioning
Note
The DSEclinical type 5198 dental simulation unit may be installed by trained
technical staff only and must be bolted to the floor in accordance with KaVo
standards.
Comply with the Technician and Installation Instructions for the assembly,
set-up and installation.
Fitting triple and multifunction handpiece
CAUTION
Damage due to missing media.
Air and water heating systems are destroyed.
▶
Check if the air and water are connected.
▶
Check the air and water supply!
▶
If possible, switch the heating off at the unit when putting into operation for
the first time or after servicing! Press the buttons carefully several times
until the media are available. Then activate heating and check its operation.
CAUTION
Damage caused by non-axial attachment.
No media coming through.
▶
Connect axially handpiece hose and handpiece!
▶
Insert the rubber dampers
③
into the gripping sleeve
②
of the handpiece.
▶
Screw hose casing
①
to valve body
④
.
27 / 100