Instructions for use DSEclinical Typ 5198
5 Operation | 5.1 Operation in general
28 / 100
5 Operation
5.1 Operation in general
5.1.1 Switching the DSEclinical type 5198 on
▶
Turn on main switch
④
.
▶
Move dentist unit
①
, assistant unit
②
, and foot control
③
into treatment po-
sition.
▶
Move the patient simulator into the treatment position.
See also:
2
Positioning the patient simulator
▶
Remove the desired instrument from the holder on the dentist unit
①
.
ð
The control automatically recognises the instrument in question
ð
(e.g. KL motor, turbine or PiezoLED).
Note
Once an instrument has been taken out of the holder, the other holders are
disabled. If a second instrument is removed from a holder, it is only recog-
nised after the first instrument is put back.
If several instruments are removed, they are enabled on the dentist's ele-
ment from left (holder 1) to right (holder 3).
Note
If the device is connected to the treatment unit with a common media sup-
ply, this treatment unit must first be turned on. Only then is the device con-
nected to the media.