Instructions for use DSEclinical Typ 5198
5 Operation | 5.5 Operating the foot switch
5.5.2 Special functions of the wireless foot control
CAUTION
Electrical power
Personal injury or damage to the wireless foot control.
▶
Do not touch the contacts of the charger connector!
The wireless foot control transmits the user activities to the treatment centre in
a wireless manner.
Rear of the wireless foot control
Item
No.
Labelling
Function
①
LED display
Status display / charge status dis-
play
②
On/Off switch
On/off switch to prevent deep dis-
charge during long periods of non-
use. The wireless foot control can
remain switched on at all times as
a matter of principle. The device
must be switched off for transport.
The battery can also be charged
when it is switched off.
③
Charge socket
Charge socket for the provided
charger (
Mat. no. 1.005.4229
).
The battery charge of the wireless foot control is indicated by the LED display
and is signaled by a tone.
Residual capacity
Foot control status
Status display /
charge status display
Beep
< 100 %
Idle state
Foot control is on
Flashes green
(approx. 2 second inter-
vals)
-
Active actuation
Flashes green
(approx. 200 millisecond
intervals)
-
< 30 %
Idle state
Foot control is on
Flashes yellow
(approx. 2 second inter-
vals)
A single brief beep when
a button is pressed.
Active actuation
Flashes yellow
(approx. 200 millisecond
intervals)
A single brief beep when
a button is pressed.
51 / 100