Instructions for use DSEclinical Typ 5198
5 Operation | 5.5 Operating the foot switch
52 / 100
Residual capacity
Foot control status
Status display /
charge status display
Beep
< 10 %
Idle state
Foot control is on
Flashes yellow
(approx. 2 second inter-
vals)
Two brief beeps when a
button is pressed.
Active actuation
Flashes yellow
(approx. 200 millisecond
intervals)
Two brief beeps when a
button is pressed.
See also:
2
5.5.12 Charging the wireless foot control, Page 56
CAUTION
Critical battery level
If the battery reaches a critical charge status, a signal is sounded every time a
function key is pressed.
▶
Always charge batteries when necessary.
▶
To ensure that the battery of the wireless foot control is always charged in
a timely manner, note the visual and acoustic signals of the wireless foot
control when starting the treatment unit.
5.5.3 Establishing a connection between the wireless
foot control and the treatment unit
Note
Only one wireless foot control per treatment centre can be registered to a RF
receiver. If another foot control was previously registered, the last foot con-
trol to be registered will be deleted with every new start of the synchronisa-
tion process.
Note
Every wireless foot control and every RF receiver has a unique address,
which can be exchanged during the synchronisation procedure. This ensures
unambiguous assignment.
The different wireless foot controls operate on different channels in order to
prevent interference during the application of several wireless foot controls.
RF receiver
①
UP key
②
DOWN key
③
"Confirmation" key