Instructions for use DSEclinical Typ 5198
5 Operation | 5.5 Operating the foot switch
5.5.7 Start and regulate instruments
Note
The foot pedal is equipped with a middle centring function, i.e., the foot pedal
always returns to levels 1 to 3 each time after it is moved to the left or right.
CAUTION
Centring in the middle is effected for the foot control by a positioning
motor.
If the positioning motor breaks down, switching from or into the middle posi-
tion using the foot control is no longer feasible. The different levels can still be
selected, but the foot pedal does not leave the middle position and cannot
switch into the middle position. The speed currently set on the wireless foot
control is always shown on the display of the treatment unit.
▶
Charge battery.
▶
If centring in the middle does not work despite the batteries being charged,
the positioning motor is defective. Have the positioning motor checked!
Note
Delays can arise after a function has been triggered if the wireless connection
of the wireless foot control is problematic.
▶
Remove the handpiece (such as turbine, motor) from the holder.
ð
The handpiece is active.
▶
Press the footpedal.
ð
The removed instrument runs at the set speed or intensity.
▶
Changing speed or intensity with the foot pedal.
ð
The left stop corresponds to the minimum speed/intensity.
ð
The right stop corresponds to the maximum speed/intensity.
5.5.8 Setting the cooling condition
▶
Remove the handpiece (such as turbine, motor) from the holder
ð
The handpiece is active.
▶
Press "Preselected spray" footswitch.
ð
The cooling level is raised each time the footswitch is pressed: no cooling -
air - spray.
ð
The cooling level is displayed on the dentist's and assistant's element.