Instructions for use DSEclinical Typ 5198
5 Operation | 5.6 Using instruments
Note
Charge the wireless foot control with the charger supplied by KaVo only.
Note
The wireless foot control charger may only be used indoors and must be pro-
tected from moisture.
▶
Connect the charger to the wireless foot control.
The charger display communicates the following:
Display
Meaning
shines green
The unit is ready
shines yellow
The battery is being charged
Shines weak green
Battery is charged
does not shine
The battery is dead, or there is a short-circuit
The battery voltage is above the tolerance range
Reversed poles
The transition from charging to full is indicated by the fluttering of the display.
5.6 Using instruments
Note
Consult the separate respective instructions for information regarding the in-
stallation, use and servicing of the individual instruments (such as the tur-
bine, INTRA LUX Motor K 200 / KL 702 / KL 703, PiezoLED, Satelec Mini LED
etc.).
5.6.1 Holder logic
All handpieces on the dentist's side are secured against simultaneous use by
holder logic. If an handpiece has been removed when the unit is switched on,
the relevant holder will not be activated until the relevant handpiece has been
replaced.
Only the withdrawn handpiece is active, i.e., any handpiece that is withdrawn
afterward is not started. Exception: MF handpiece (parallel operation is possible
here).
5.6.2 Using suction hoses
▶
Remove the spray mist suction device or saliva ejector from the holder.
ð
The spray mist suction device or saliva ejector automatically turns on, and
then when it is placed in the holder, it turns off.
The suction flow of the saliva ejector or spray mist suction device can be re-
duced or blocked with the slide valves integrated in the handpieces.
57 / 100