Instructions for use DSEclinical Typ 5198
5 Operation | 5.6 Using instruments
▶
Install a new high pressure lamp (
Mat. no. 1.002.2928
).
Replacing the KaVo MULTI LED lamp
Note
The KaVo MULTI LED bulb is a semiconductor element and may only be oper-
ated with direct current. To ensure proper function, the poles need to be in-
serted correctly.
▶
Push the holder
③
forward and pull the defective KaVo MULTI LED lamp
②
out of the socket.
▶
Insert new Kavo MULTI LED lamp (
Mat. no. 1.007.5372
).
The following may happen after you turn on the KaVo MULTI LED lamp:
▪ Case 1: KaVo MULTI LED lamp is on.
▪ Case 2: KaVo MULTI LED lamp is faint.
- Increase the cold light intensity on the unit until the desired light intensity
is reached.
▪ Case 3: KaVo MULTI LED lamp is red or off.
- Take KaVo MULTI LED lamp out of its socket as described above and re-in-
sert it after rotating it 180° about its axis.
5.6.5 Using KaVo Poly One
Warning
Danger from optical radiation.
Eyes can be blinded or damaged by LED of risk group 2 in accordance with
IEC 62471 (blue light).
▶
Do not gaze directly into the light exit aperture of the fibre-optic cable or
into the lamp.
See also:
2
Instructions for use of KaVo Poly One
63 / 100