Instructions for use DSEclinical Typ 5198
5 Operation | 5.7 Using the KL702 / KL 703 in ENDO mode (optional accessory)
▶
To adjust the intensity, move the "Instruments" foot pedal to the side.
5.7 Using the KL702 / KL 703 in ENDO mode (optional
accessory)
5.7.1 General information
Note
The Endo drive can only be operated with the INTRA LUX KL 703 LED or IN-
TRA LUX KL 702 motors.
INTRA LUX KL 703 LED
INTRAmatic LUX KL 702
CAUTION
Use of impermissible filing systems.
Do not use impermissible filing systems which can damage the product or
cause personal injury.
▶
Only use approved NiTi file system with a conicity >2% that are suitable for
rotary reprocessing.
▶
Only use files with shafts in conformance with DIN EN ISO 1797, DIN EN
ISO 3630-1 and DIN EN ISO 3630-2 having a shaft diameter of 2.334 to
2.350 mm
▶
Follow manufacturer's instructions (mode of operation, speed, torque
levels, torsion resistance, etc.), and use the files according to their intended
use.
CAUTION
Use of damaged files.
Injury to the patient or damage to the medical device.
▶
Before preparing each root canal, insert a dental dam for safety reasons.
▶
Before each use, the files must be checked for possible signs of
material fatigue, deformation or excessive stress and if such signs appear,
they must be replaced.
CAUTION
Incorrect transmission ratio.
Damage from incorrect speed / incorrect torque.
▶
Only use KaVo 1:1 reducing shanks 20LH or 20LP or MASTERmatic LUX
M20 L with 1:1 INTRA LUX head L68 B (
Mat. no. 1.008.1834
) or 3:1 IN-
TRA head L66 B (
Mat. no. 1.008.1831
).
65 / 100