Instructions for use DSEclinical Typ 5198
5 Operation | 5.7 Using the KL702 / KL 703 in ENDO mode (optional accessory)
66 / 100
CAUTION
Excessive torque.
Injury or damage to instruments.
▶
Use root canal instruments in ENDO mode only.
Technical specifications for the KL 703 LED / KL 702 in
ENDO mode
Note
The technical specifications apply to the KL 703 LED / KL 702 in ENDO mode.
Speed range
200 to 3,200 min
-1
Maximal torque
2.5 Ncm
Mode
Note
30 seconds operating time/9 minutes pause is the potential load threshold of
the motor (full load at maximum speed).
In practice, pulse loads lasting seconds or pause times lasting seconds or
minutes are realistic given that the maximum possible motor current is not nor-
mally reached. This equates to the dentist's normal way of working.
5.7.2 Open ENDO mode
▶
Remove the endomotor INTRA LUX Motor KL 703 / KL 702 from the holder.
▶
Press the "Additional motor drives" key.
ð
The display switches to the "ENDO" menu.
Note
Before using the endomotor, always check the speed and transfer ratio.
The Endo mode is left as soon as the INTRA LUX KL 703 LED / KL 702 endomo-
tor is returned to the holder. The Endo mode is activated automatically once the
endomotor is removed, providing that the Endo mode was previously ended by
putting the endomotor back in place.
Note
The device does not automatically start when endo mode was left by pressing
the "Motorised auxiliary drives" button, or if endo mode has never been ac-
tivated since the last time the unit was turned on.
A service technician can deactivate the automated start.