Instructions for use DSEclinical Typ 5198
5 Operation | 5.7 Using the KL702 / KL 703 in ENDO mode (optional accessory)
Parameters in the display
①
Direction of motor rotation
②
Torque mode
③
Transmission factor
④
ENDO Mode
⑤
Speed
⑥
Parameter memory
⑦
Torque
CAUTION
Incorrectly set parameters.
Injury or property damage from incorrect input values.
▶
Check all input values before use (e.g. transmission ratio, transmission
factor, torque, etc.).
5.7.3 Change settings in the option menu
▶
Press the "Next" key to switch to the option menu.
ð
The last used menu is displayed.
The following settings can be changed in the option menu:
Display
Function
Option: 1. Transmission factor
Set the transmission factor to 1:1 or
3:1
Option: 2. Display of torque
Set the torque display in Ncm or in %
1:1 transmission: 100% = 2.5 Ncm
3:1 transmission: 100% = 8 Ncm
Option: 3. Autorev./Fwd. time
In Autorev./Fwd time mode, you can
set the time (1 to 10 seconds) in which
the motor automatically starts rotating
to the right which makes it unneces-
sary to stop with the foot control.
Options 1, 2 or 3 can be selected by pressing the "Back" and "Forward" keys.
You can change the option parameters with the "Increase value" and "De-
crease value" keys.
▶
Press the "Enter" key to leave the option menu.
ð
The changed parameters are saved.
5.7.4 Set parameters
There are six parameter memories (P1 to P6).
67 / 100