5 Operation | 5.7 Using the KL702 / KL 703 in ENDO mode (optional accessory)
68 / 100
The following parameters can be changed:
▪ Speed
▪ Torque
▪ Torque mode
Select the parameter memory
▶
Press "Programme" button in order to call-up the desired parameter
memory location (P1 to P6).
or
▶
Press the "SP/Blown air" foot button.
ð
Each time the button is pressed, the parameter memory location advances
by one step (P1 - P2 - P3 - … - P6 - P1)
Change and save parameters
▶
Press the "Program" key to open the desired parameter memory (P1 to P6).
▶
Press the key for "Select parameters" to select the desired parameter.
ð
The cursor flashes on the parameter to be changed.
▶
Press the "UP" or "DOWN" key to change the selected parameter.
Note
If you hold down the "Up" and "Down" keys repeat automatically.
▶
To save the parameters, press the "Program" key for two seconds until you
hear the signal.
ð
The changed parameters are saved in the selected parameter memory.
Note
You can save the parameters each time you set a new one, or after setting
all the parameters.
Setting the speed
The speed can be adjusted in steps of 10 from 100 rpm to 500 rpm, in steps of
50 from 500 rpm to 1000 rpm, and in steps of 100 from 1000 rpm to 6000
rpm.
▶
Press the "DOWN" key to reduce the speed.
or
▶
Press the "Up" key to increase the speed.
ð
The speed is shown on the display and is effective immediately.