5 Operation | 5.7 Using the KL702 / KL 703 in ENDO mode (optional accessory)
Save the parameters in programs 1 to 6 using the "Program" key (press for 2
seconds). You can save after setting one parameter, or after setting all para-
meters. Saving is acknowledged with a tone.
Setting the torque
The torque is limited to the set value.
Note
The ENDO warning signal sounds when you reach 90% of the set torque.
1:1 transmission ratio
The torque can be set in 0.05 Ncm steps in the range from 0.15 Ncm to 2.5
Ncm, or in 2% steps in the range from 1% to 100%.
3:1 transmission ratio
The torque can be set in 0.05 Ncm steps in the range from 0.4 Ncm to 8 Ncm,
or in 1% steps in the range from 1% to 100%.
See also:
2
▶
Press the "Down" key to reduce the torque.
or
▶
Press the "Up" key to increase the torque.
ð
The torque is shown on the display and is effective immediately.
Save the parameters in programs 1 to 6 using the "Program" key (press for 2
seconds). You can save after setting one parameter, or after setting all para-
meters. Saving is acknowledged with a tone.
Set torque mode
Three different torque modes are available:
▪ Autoreverse
▪ Torque Control only
▪ Autorev / Forward
▶
Press the "Up" or "Down" key to select the desired torque mode.
ð
The torque mode is shown on the display and is effective immediately.
Save the parameters in programs 1 to 6 using the "Program" key (press for 2
seconds). You can save after setting one parameter, or after setting all para-
meters. Saving is acknowledged with a tone.
Torque mode Torque Control only
▶
Press the foot pedal.
69 / 100