Instructions for use DSEclinical Typ 5198
6 Service and Care | 6.2 Care
74 / 100
Cleaning and disinfection of the handpiece holder
NOTICE
Damage caused by treatment in a washer disinfector.
▶
Do not treat the instrument holder in a washer disinfector.
▶
Pull the snap-in buttons outward evenly and carefully remove the instru-
ment holder.
▶
Clean the instrument holder under running water.
▶
Wipe-disinfect the instrument holder.
Clean the lamp
See also:
2
Instructions for Use of the respective lamp
6.2.1 Setting up the 3-function and multifunction
handpiece
Note
Wet the O-rings of the gripping sleeves after sterilisation using silicone
grease (
Mat. no. 1.000.6403
) or KaVo Rota Spray 2 (
Mat. no.
0.411.7510
). After replacing the cannula, the air channel must be blown out
before the start of treatment to ensure that the air exiting during treatment
is dry.