Instructions for use DSEclinical Typ 5198
6 Service and Care | 6.2 Care
Cleaning the triple-function and multifunctional
handpiece
Automatic cleaning
Not applicable.
NOTICE
Damage caused by improper cleaning/disinfection.
▶
Do not wash the gripping sleeve and the cannula in the washer disinfector.
▶
Do not sterilise the gripping sleeve and the cannula in the hot-air steriliser.
▶
Do not place the gripping sleeve and the cannula in a disinfectant or ultra-
sonic bath.
Manual cleaning
Note
Combine manual cleaning with disinfection.
Cleaning the interior
▶
Leave the gripping sleeve and cannula on the instrument, and activate the
air and water for 60 seconds.
External cleaning
CAUTION
Danger of injury if the valve body is hot (heating elements, high-pres-
sure lamp).
Risk of burn injury.
▶
Switch off the unit at the main switch.
▶
Let the handpiece cool after long use.
Accessories required:
▪ Tap water 30
o
C ± 5
o
C
▪ Brush, e.g. medium-hard toothbrush
▶
Disassemble the gripping sleeve and cannula.
▶
Detach the key part from the gripping sleeve.
▶
Clean the key part, gripping sleeve and cannula by brushing them under
running water (at least tap water quality).
▶
Immediately remove contamination from materials used in the dental prac-
tice (impression materials, caustic chemicals).
▶
Carefully wipe off contamination from the cannula tip with a soft Q-tip or
soft cloth and water.
75 / 100