6 Service and Care | 6.2 Care
76 / 100
NOTICE
Damage caused by using a nozzle needle that is too long.
Damage to the internal hoses of the cannula.
▶
Use a short nozzle needle (
Mat. no. 1.004.4986
) to clean the media exit
holes.
▶
Carefully remove contamination from the media exit ports using the short
nozzle cleaning needle.
Disinfection of the 3-function and multifunction
handpiece
Manual disinfection
External disinfection
▶
Wipe-disinfect the surfaces with a soft cloth.
Automatic disinfection
Not applicable.
NOTICE
Damage caused by improper cleaning/disinfection.
▶
Do not wash the gripping sleeve and the cannula in the washer disinfector.
▶
Do not sterilise the gripping sleeve and the cannula in the hot-air steriliser.
▶
Do not place the gripping sleeve and the cannula in a disinfectant or ultra-
sonic bath.
Drying the 3-function and multifunction handpiece
Manual drying
▶
Blow off the outside and inside with compressed air until water drops are no
longer visible.
Automatic drying
Not applicable.
Care of the 3-function and multifunctional handpiece
Requirement
The multifunctional handpiece has been reprocessed.
▶
Apply KaVo silicone grease (
Mat. no. 1.000.6403
) or KaVo Rota Spray 2
(
Mat. no. 0.411.7510
) to the O-rings of the joint between the gripping
sleeve and cannula. Use cotton buds for this purpose.