Instructions for use DSEclinical Typ 5198
6 Service and Care | 6.2 Care
Sterilising the 3-function and multifunction handpiece
Note
The grip sleeve and cannula can be sterilised.
▶
Hold the 3-way or multifunctional handpiece at the gripping sleeve and take
off the cannula with a slight twisting motion.
▶
Pull the gripping sleeve off the valve body.
Sterilization in a steam sterilizer in compliance with DIN
EN 13060
▶
Weld the cannula and grip sleeve in sterile goods packaging.
▶
Sterilise with a triple fractionated vacuum (B class steriliser); hold for 4
minutes; 134
°
C ± 1 °C; overpressure: 2.13 bar.
NOTICE
Damage caused by over-heating.
Damage to the gripping sleeve.
▶
Immediately remove the parts from the steriliser after the sterilisation cycle
is completed.
Storage
▶
Reprocessed products should be stored protected from dust with minimum
exposure to germs in a dry, dark and cool place.