6 Service and Care | 6.2 Care
78 / 100
6.2.2 Automatic hygiene functions of the water
conducting system
Note
Intensive germ reduction kit and water block DVGW required.
CAUTION
Health damage due to germ formation.
Infection hazard.
▶
Before starting, rinse all the water drain lines with no handpieces attached.
▶
Before initial start-up and after downtimes (weekends, holidays, vacations,
etc.), rinse or purge the air and water lines with air.
▶
Carry out an intensive germ reduction.
The water-conducting system includes all water ducts of the treatment centre
to which the patient may be exposed. This includes the instruments on the
dentist element.
During normal operation of the treatment centre, the continuous germ reduc-
tion function secures continuous germ reduction in the water conducting sys-
tems due to the automatic dosing of the germ reduction agent,
KaVo OXYGENAL 6.
Rinsing programme and intensive germ reduction facilitate the rinsing and germ
reduction of the water-supplying system. This counteracts the formation of mi-
cro-organisms at times of stagnation and ensures a hygienic water supply.
Rinsing programme
The rinsing programme 02 provides for automatic rinsing of the water-conduct-
ing system and all instruments for 2 minutes each (hydrogen peroxide concen-
tration 0.02 %). In the short rinsing programme 01, all instruments on the
dentist unit and assistant unit are rinsed for 20 seconds each.
Procedure of the rinsing programme:
▪ Before you start working (rinsing programme 02)
▪ After each patient and in the evening (rinsing programme 01)
▶
Briefly press the "Intensive germ reduction" button.
Upper LED flashes.