Instructions for use DSEclinical Typ 5198
6 Service and Care | 6.2 Care
Preparation for rinsing programme and intensive germ
reduction
Placing the handpieces in the hygiene centre
▶
Open the door of the hygiene centre.
The cannulas of the three-function and multifunctional handpieces must be re-
moved for the cleaning programmes.
▶
Grasp the three-function or multifunctional handpiece on the adapter
between hose and handpiece and take off the cannula and the gripping
sleeve.
▶
Place the three-function or multifunctional handpieces without cannulas or
gripping sleeves in the middle holders of the hygiene centre. Make sure to
place triple function handpieces in the grey holder and multifunctional hand-
pieces in the black holder of the hygiene centre.
▶
Place the instrument hoses of the dentist and assistant element in the
places provided in the hygiene centre.
Note
All water regulation mechanisms for the motor, instruments, and multiflex
couplings must be open.
79 / 100