6 Service and Care | 6.2 Care
80 / 100
Placing the COMFORTbase in the hygiene centre
Note
A seat adapter for the germ reduction attachment is available for the COM-
FORTbase.
▶
For the cleaning programmes, press the seat adapter into one of the six
holders of the hygiene centre.
▶
Place the COMFORTbase in the seat adapter.
Carry out the rinsing programme
Set rinsing time
▶
Press the „decrease value“ or „increase value“ keys to select rining program
01 „Programm vorbereiten: 01“ or rinsing program 02 „Programm
vorbereiten: 02“.
ð
The standard value for rinsing program 01 is 00:20 minutes per instrument.
ð
The standard value for rinsing program 02 is 2:00 minutes per instrument.
▶
Press the foot control to display the time for rinsing the instrument hoses
and the three-function or multifunctional handpiece.
With the foot pedal of the foot control being pressed down, the rinsing time for
the respective rinsing programme can be set in the range from 00:01 to 4:00
minutes using the "Decrease value" and "Increase value" buttons.
The rinsing program proceeds in four steps. For rinsing programme 01, the
cycle starts in step 2, and for rinsing programme 02, the cycle starts in step 4.
Step
Description
4.0
Fresh water supply
Duration: 60 seconds
3.0
Tumbler rinsing
Duration: 20 seconds
2.0
Rinsing of the handpieces
The rinsing time can be set variably for each handpiece.
Standard setting:
▪ Rinsing programme 01 = 20 seconds per instrument
▪ Rinsing programme 02 = 2 minutes per instrument
1.0
Rinsing the triple-function and multifunction handpiece
Set rinsing time of rinsing programme 01 or 02
0.1
Waiting position for setting down the handpieces