Instructions for use DSEclinical Typ 5198
6 Service and Care | 6.2 Care
Start rinsing programme
Note
The rinsing program can be started immediately or the next morning.
▶
Briefly press the "Intensive germ reduction" button to immediately start the
process.
ð
The selected rinse program starts.
or
▶
Turn off the device and do not restart it until the next morning.
ð
The cycle is started automatically. The rinsing time follows rinsing program
02.
Note
If the device is not used the entire day (i.e., the instruments are not moun-
ted), the rinsing program automatically starts the next morning when the
device is turned on.
The rinsing programme can be stopped at any time by pressing the "intensive
germ reduction" key.
If the device is turned off and then on again, the rinsing program also stops.
▶
After completion of the rinsing program, set down the instruments.
ð
The LED of the "intensive germ reduction" button is not lit up.
Intensive germ reduction
Note
Observe the instructions for use for the instruments.
Intensive germ reduction prevents the formation of microorganisms during ex-
tented periods when the equipment is out of use. During intensive germ reduc-
tion, the afferent water ducts are automatically filled with a higher concentra-
tion of KaVo OXYGENAL 6. The intensive germ reduction program lasts ca. 45
minutes. The increased hydrogen peroxide concentration acts in the system for
at least 30 minutes.
An additional intensive germ reduction must be carried out during extended
downtimes (> 4 weeks).
The need for intensive germ reduction is automatically displayed in the top LED
of the "intensive germ reduction" button at initial start-up and after downtimes
(weekend, holidays, vacation, etc.).
If a sound is emitted every 10 seconds and the lower LED of the "Intensive
germ reduction" button flashes, the OXYGENAL 6 bottle must be replaced.
See also:
2
Replace the OXYGENAL 6 bottle, Page 000
Note
If the amalgam separator malfunctions during intensive germ reduction, the
entire process is stopped. The bottom LED of the "HYDROclean" button
flashes.
81 / 100