Instructions for use DSEclinical Typ 5198
6 Service and Care | 6.2 Care
82 / 100
Intensive germ reduction must be carried out:
▪ At initial start-up
▪ Before, during and after prolonged downtimes (more than 4 weeks)
▪ Over the weekend
▪ During vacations
Start intensive germ reduction
▶
Press the "Intensive germ reduction" button for more than 4 seconds until
you hear an acoustic signal and the upper LED of the "intensive germ re-
duction" button flashes. The display switches to the intensive germ reduc-
tion menu.
▶
Perform the preparations for intensive germ reduction. (Place instruments
and suction hoses in the hygiene centre.)
See also:
2
6.2.2.1.1 Preparation for rinsing programme and intensive germ reduction,
Page 79
▶
Press the "Intensive germ reduction" button.
ð
This starts up the intensive germ reduction, which proceeds through several
stages. The process can be followed on the display. The major steps are lis-
ted below.
Step
Description
7.0
Germ reduction of the triple-function and multifunctional handpiece
and the spittoon bowl
Duration: 60 seconds
6.0
Germ reduction of the handpieces
Duration: 30 seconds each
5.0
Germ reduction of the tumbler
Duration: 40 seconds
4.0
Dwell time (acoustic signal, switch off is possible)
Duration: 30 minutes
3.0
Rinsing of the tumbler
Duration: 40 seconds
2.0
Rinsing of the handpieces
Duration: 30 seconds
1.0
Rinsing of the triple-function and multifunctional handpiece
Duration: 30 seconds
0.1
Waiting position for setting down the handpieces
Intensive germ reduction is completed when "01 intensive germ reduction
completed" appears on the display.