6 Service and Care | 6.2 Care
▶
Place all instruments in the hygiene centre and close the flap of the hygiene
centre.
Interrupt intensive germ reduction during the dwell
time
So that no one will have to remain with the device throughout the entire intens-
ive germ reduction process, intensive germ reduction can be interrupted in pro-
gram step 4 (e.g. on the last workday before the weekend). Program step 4 is
the dwell time, in which the microorganisms are killed by an increased
OXYGENAL 6 concentration.
Requirement
intensive germ reduction is in program step 4.
Requirement
You hear a beep for 4 seconds.
▶
Turn the device off.
When the device is switched on after 30 minutes of dwell time, the remaining
program steps run immediately and automatically.
If the device is turned on before the 30 minute period is elapsed, the dwell
times finishes first before the remaining steps of the intensive germ reduction
program run.
Skip exposure phase
This function is required when the treatment unit needs to be used (for ex-
ample in an emergency, at hospitals, etc.) even though the germ reduction pro-
cedure has already started.
See also:
2
6.2.2.1.2 Workflow of the rinsing programme, Page 80
Intensive germ reduction is in step 4:
▶
Press and hold down the "Intensive germ reduction" button and the "Hydro-
colloid" button simultaneously.
▶
Simultaneously press the foot pedal on the foot control.
ð
The intensive germ reduction switches to step 3.7.
83 / 100