7 Safety check according to VDE 0701 / 0702
The applicable national legal regulations must be observed during the use of the
device, in particular:
▪ the applicable occupational health and safety regulations.
▪ the applicable accident prevention regulations.
The overarching guidelines and/or national laws, national regulations and the
rules of technology applicable to this product for start-up and use of the KaVo
product for the intended purpose are to be applied and complied with.
Compliance with the requirements of VDE 0701-0702 concerning product safety
upon start-up, after changes and repair as well as in recurrent tests is required.
Caution: Electrical safety checks have to be carried out by electrical engineering
professionals or by individuals trained in electrical engineering and supervised
by an electrical engineering professional.
An electrical engineering professional is defined to be anyone whose training,
knowledge and experience as well as knowledge of the pertinent regulations al-
low him or her to assess the imminent work and recognise possible hazards. It
is at the discretion of the electrical engineering professional in charge to decide
whether nor not further individual tests are required in order to attain the
safety target.
If safety deficiencies or damage is/are detected during the recurrent test, the
testing process must be discontinued and the equipment must be labelled ac-
cordingly and banned from further use or submitted for repair of the defect.