Instructions for use DSEclinical Typ 5198
8 Troubleshooting
8 Troubleshooting
Malfunction
Cause
Remedy
The three-function
handpiece or multifunc-
tional handpiece is not
working properly.
Water pressure or air
pressure are set incor-
rectly.
▶
Notify a service technician.
Shut-off valve for water
or compressed air
closed.
▶
Notify a service technician.
Nozzle blocked.
▶
Clean the nozzle.
▶
Replace nozzle according to need.
K 200 / KL 702 / KL
703 motor handpiece
cannot be started.
Mains switch not at "I"
ON setting.
▶
Push the mains switch to the "I" ON setting.
Patient simulator/back-
rest cannot be moved/
adjusted.
Air supply is interrup-
ted.
▶
Notify a service technician.
Turbine not running.
Air supply broken.
▶
Notify a service technician.
Suction power of saliva
extractor not adequate.
Spray mist extractor
not closed.
▶
Close spray mist extractor.
No voltage supply
Main switch is off
▶
Turn on main switch.
No voltage supply
Main fuse is blown.
▶
Unplug the unit from the mains.
▶
Change the microfuse T 6.3 H (
Mat. no.
0.223.2783
).
The main fuse is next to the main switch.
Cold light not working
Cold light not preselec-
ted.
▶
Preselect cold light.
High pressure lamp
faulty.
▶
Renew high pressure lamp.
No spray in the instru-
ments
No spray preselected.
▶
Preselect spray.
Spray regulating device
on instruments closed.
▶
Open spray regulating control.
Main valve closed.
▶
Open main valve.
Turbine making loud
running noises.
Turbine wheel faulty.
▶
Replace turbine wheels.
▶
Follow turbine instructions.
Satelec Mini LED /
KaVo Poly One does not
work.
Also refer to:
Instruc-
tions for Use of the
Satelec Mini LED /
KaVo Poly One
93 / 100