Instructions for use DSEclinical Typ 5198
8 Troubleshooting
94 / 100
Malfunction
Cause
Remedy
Spray at instruments
works poorly.
Water pressure or air
pressure are set incor-
rectly.
▶
Notify a service technician.
Calcified/dirty spray
nozzles.
▶
Clean spray nozzles in accordance with the en-
closed instrument instructions.
Leaks on instruments
O-rings damaged at
multiflex/motor coup-
ling.
▶
Notify a service technician.
Water in return air fil-
ter.
O-rings damaged at
multiflex coupling.
▶
Notify a service technician.
Suction hoses not
working.
Slides on the bodies
are closed.
▶
Open the slide valve.
Sieves in suction con-
nector blocked.
▶
Notify a service technician.
Suction machine not
running.
▶
Check connectors of suction machine DSE unit.
Malfunction
Cause
Remedy
Display shows: ID 33
Chair operation is
stopped.
▶
Call a service technician.
Display shows: ID 56
Wireless foor control
battery is empty.
▶
Charge battery.
Display shows: ID 64
Water is shut off.
▶
Turn water on.
Water works leaks
strongly.
Water works malfunc-
tion
▶
Call a technician.
Display shows: ID 65
Safety switch of bowl
suction has been
reached.
▶
Turn external suction on.
▶
Notify a service technician.
Display shows: ID 66
Amalgam separator
malfunction
▶
Remedy the malfunction.
See also:
2
Instructions for Use of amalgam separator
Display shows: ID 67
The Oxygenal container
is empty.
▶
Refill the Oxygenal container.
See also:
2
Replenishing and checking the disinfectant
Display shows: ID 68
Call for service
▶
Have a service performed.
▶
Call a technician.