Instructions for use DSEclinical Typ 5198
8 Troubleshooting
Malfunction
Cause
Remedy
Display shows: ID 69
An intensive germ re-
duction must be done.
▶
Carry out an intensive germ reduction.
See also:
2
Intensive germ reduction
Display shows: ID 72
Dekaseptol bottle.
▶
Insert the DEKASEPTOL bottle.
See also:
2
Care instructions
Display shows: ID 74
Indicator if the DEKA-
SEPTOL bottle is not
full after 600 s (10
min).
Centramat empty.
▶
Notify the service centre.
Display shows: ID 75
Centramat too full.
▶
Notify the service centre.
Display shows: ID XX
This error is not de-
scribed in this chapter.
▶
Call a technician.
Display shows: CAN fail Internal communication
error.
▶
Turn the unit off and on again, consult a ser-
vice technician according to need.
Display shows: System
State
Unit does not work.
▶
Call a service technician.