Instructions for use DSEclinical Typ 5198
6 Service and Care | 6.2 Care
84 / 100
ð
The automatic rinsing process starts up.
Intensive germ reduction during extended periods of
non-use (> 4 weeks)
During extended periods of non-use, e.g. semester breaks at university, intens-
ive germ reduction must be carried out during the time of non-use. The germ
reduction fluid remains inside the system during the period of non-use.
Before the period of non-use
▶
Start the intensive germ reduction and stop it in step 4 after the signal tone
during the dwell time of 30 minutes.
▶
To prevent damage due to hoses getting kinked, remove hoses from the
germ reduction attachment and place them in the dentist support.
▶
Keep the germ reduction attachment in the spittoon bowl.
▶
Cover the treatment unit during extended periods of non-use.
After the period of non-use
▶
Turn on the central compressor.
▶
Switch on the water supply for the treatment unit after the system pressure
is achieved.
▶
Replace the instrument hoses of the dentist and assistant units in the places
provided in the germ reduction attachment.
▶
Turn the treatment unit on.
ð
The unit automatically starts the remaining steps of the unfinished cycle.
▶
Allow the cycle to complete and then carry out a second full intensive germ
reduction.
ð
The unit is now ready for use.
Additional intensive germ reduction during period of
non-use
KaVo recommends carrying out an additional intensive germ reduction at uni-
versities during semester breaks. This flushes the parts of the unit and replaces
the germ reduction liquid.
6.2.3 Reprocessing and automatic hygiene functions of
the suction system
HYDROclean function
The HYDROclean function is an automatic programme for rinsing the drainage
and separation system (limited rinsing). The water is aspirated from the
Aquamat through the filter housing to the separation system.
The HYDROclean function prevents the coagulation and adhesion of proteins in-
side the drainage system and improves the discharge of fine particles.
Carrying out the HYDROclean function:
▪ After each treatment