Instructions for use KaVo EDUnet ULTRA HD
4 Assembly | 4.2 Connect the bus cable
4.2 Connect the bus cable
All participant boxes are connected to each other by the bus cable in a ring system.
The physical arrangement of the participant boxes in the bus depends on their code/
arrangement on the room keyboard.
NOTICE
Malfunction due to inappropriate cables.
Image transmission error.
▶ Always use the manufacturer-released patch cables included in the delivery.
▶ If you use other cables, have the cables tested by the manufacturer.
▶ Use only bus cables with a length of no more than 30 m.
▶ Do not use UTP cables.
NOTICE
Product damage due to kinking or mechanical pulling.
Irreversible breakage of the electrical lines in the bus cable.
▶ Always run the bus cables with sufficient play.
▶ Do not kink bus cables.
▶ Do not pull on bus cables.
Note
The maximal length of the bus cable between two participants is 30 m. If the ar-
rangement of the participants is complicated, the ring network can be split into
"branches" with a switch.