Instructions for use ELECTROmatic M/C and PM/PC
5 Commissioning | 5.3 Measure cooling air quantity at motor coupling
52 / 80
5.3 Measure cooling air quantity at motor coupling
▶
Place the airflow measuring tube
①
(
Mat. no. 0.411.4441
) on the motor.
▶
Press the foot control to start the motor.
▶
Adapt the system pressure of the treatment unit appropriately such that the
cooling airflow is 6 to 9 Nl/min (upper edge of sphere
②
).
▶
Comply with the system pressure limits from the technical specifications for
media.
See also:
2
3.6 Technical specifications of the ELECTROmatic, Page 25
5.4 Making the device settings
▶
Press both the plus and minus key for 2 seconds to start the "Settings"
menu.