Instructions for use ELECTROmatic M/C and PM/PC
10 Troubleshooting
72 / 80
Malfunction
Cause
Remedy
Event E2
Foot control is being
pressed during start-
up.
▶
Do not press the foot control when you switch
the unit on.
Event E3
No motor is connected.
▶
Connect the motor.
Event E4
Motor is blocked.
▶
Relieve the motor of load, stop and start on
foot control.
Event E5
Automatic light and
motor shut-off during
continuous operation of
the motor.
▶
Comply with defined operating mode.
Event E8
Saved data or settings
have been reset to
starting value.
▶
Confirm message and check or correct the pro-
gram settings.
▶
If error persists, notify the service engineer.
Event E12
No connection to mo-
tor.
▶
Check/restore correct connection.
▶
Replace the motor cord.
▶
If error persists, notify the service engineer.
Event E14
Motor overload.
▶
Relieve the motor of load, stop and start on
foot control.
All other events
Internal system error.
▶
Turn the unit off and on.
▶
If error persists, notify the service engineer.