Instructions for use ELECTROmatic M/C and PM/PC
6 Operation | 6.4 Changing the speed setting
58 / 80
6.4 Changing the speed setting
Note
In the ELECTROmatic M/C without display, the speed can be changed only
with the foot control of the treatment unit. If a foot control without steps is
actuated, the instrument is operated at the maximum motor speed of 40,000
min
-1
(rpm).
To make sure that a certain speed is not exceeded and/or is maintained in
the ELECTROmatic M/C without display, a reducing contra-angle handpiece
needs to be used (e.g. EXPERTmatic LUX contra-angle handpiece E15 L,
MASTERmatic LUX M07 L or MASTERmatic LUX M29 L).
The user can set 3 speed memories and invoke by pressing the keys, "1", "2"
or "3".
The following factory settings a pre-selected as defaults for the speed memo-
ries:
Speed memory
INTRA LUX KL 703 LED
[min
-1
/rpm]
COMFORTdrive 200 XDR
[min
-1
/rpm]
1
3,000
20,000
2
20,000
160,000
3
40,000
200,000
▶
Press the "1", "2" or "3" key.
ð
The display shows the set speed (speed min
-1
x 1,000).
▶
Press the Plus or Minus key to change the saved value.
ð
The display shows the speed.
▶
Press the "1", "2" or "3" key for more than 3 seconds to save the value in
the respective speed memory.
ð
Speed memory "1", "2" or "3" flashes and an audio signal is issued.
ð
The set value is being saved.
6.5 Changing the direction of rotation
▶
Press the "Direction of motor rotation" key.
ð
The direction of motor rotation changes.
ð
The symbol on the display indicates the direction of motor rotation.
6.6 Protection function SAFEdrive (PM/PC only)
WARNING
Use of incorrect straight and contra-angle handpieces.
Risk of burn injury or overheating.
▶
Use only original KaVo high-speed handpieces of the 25L, E25, M25, M05,
25LP, 25LPA, 25LPR, 25LCA series or the COMFORTdrive motorized contra-
angle handpiece.