Care instructions ESTETICA E70 Vision / E80 Vision
2 Setup methods according to ISO 17664 | 2.4 Preparation of the dentist unit
Sterilize the instrument holder
NOTICE
Damage caused by sterilisation.
▶
Do not sterilise the instrument holder.
2.4.3 Reprocessing of the swing-arm S table (E70 Vision
only)
Note
The swing-arm S-table (
Mat. no. 10074755)
) can be removed to facilitate
cleaning.
Cleaning and disinfection of the instrument arm S table
NOTICE
Damage caused by treatment in a washer disinfector.
▶
Do not sterilise the instrument arm S table in a washer disinfector.
Note
No cleaning programmes can be run when the swing-arm is removed.
▶
Locking the screen.
See also:
2
Locking the touchscreen, Page 9
▶
Take the handpieces off the hoses.
▶
Take the roller out of the swing-arm and then place the hose in the hygiene
centre.
▶
Pull-off the swing-arm in an upward direction.
▶
Clean the swing-arm and roller under running water.
13 / 48