Instructions for use ESTETICA E70 Vision / E80 Vision
4 Operation | 4.13 Using instruments
▶
Simultaneously press the air button
①
and water button
④
and continu-
ously increase or decrease the exiting spray by applying more or less pres-
sure on the two buttons.
Removing the cannulas
▶
Hold the 3-way or multifunctional handpiece at the gripping sleeve and take
off the cannula with a slight twisting motion.
4.13.4 Using the multifunctional handpiece
CAUTION
Risk of injury from touching the cheek with the handpiece.
Irritation of the mucosa.
▶
Rotate the cannula of the handpiece into an operating position where there
is no contact of the mucosa.
CAUTION
Cannula that is worn or not locked into place.
Injury from swallowing the cannula.
▶
Before each treatment, ensure that the cannula is locked into place and
firmly seated.
▶
Use original KaVo cannulas only.
▶
Only use reliable cannulas showing no damage.
CAUTION
Insufficient clearance between cannula and surface of gums or gin-
giva.
Injury hazard.
▶
Adhere to a minimum clearance of 10 mm between cannula and surface of
gums or gingiva.
NOTICE
Damage due to missing media.
Air and water heating systems are destroyed.
▶
Check if the air and water are connected.
▶
Check the air and water supply!
▶
If possible, switch the heating off at the unit when putting into operation for
the first time or after servicing! Press the buttons carefully several times
until the media are available. Then activate heating and check its operation.
109 / 178