Instructions for use ESTETICA E70 Vision / E80 Vision
4 Operation | 4.15 Using the SL600 surgical motor (optional accessory)
128 / 178
▶
Hold down the "Blown air" foot-operated button for a while.
Setting and saving parameters
▶
Selecting the working step to be changed.
The following parameters can be changed:
▪ Maximum torque
▪ Programmed speed
▪ Coolant flow
▪ Transmission ratio
▪ Direction of motor rotation
▪ Activity
▶
Set the parameters as in the instrument settings.
The values shown for a programme step are default values that you can use to
start work immediately. These can all be changed and thus adapted to your in-
dividual work technique.
Changed values can be saved and are then available for the next use.
CAUTION
Incorrect application
Injury hazard.
▶
Always check the values prior to application.
Recommended programming when placing multiple
implants consecutively
Note
Default values for the respective transfer ratios can be selected with the
sliders.
Activity
Step
Icon
Speed
[rpm]
Torque
[Ncm]
Transmis-
sion ratio
Pilot drilling 1
200 – 2,000
500 (D)
5 – 20
10 (D)
1:1
16:1
27:1
20:1 (D)