Instructions for use ESTETICA E70 Vision / E80 Vision
4 Operation | 4.16 Use pump for physiological saline solution (optional accessory)
138 / 178
Note
Make sure to place the pump hose in the pump appropriately such that the
pump hose does not get clamped or pinched by the lock. Route all hoses re-
laxed and without tension.
▶
Stick the puncture needle
⑥
into the coolant container and hook-in the
coolant container on the bottle holder.
▶
Check the sealing and firm seating of the puncture needle
⑥
. Prevent fluid
from leaking above the device.
▶
If you use a glass bottle, open the ventilation on the puncture needle
⑥
.
▶
If you use a bag, keep the ventilation on the puncture needle
⑥
closed.
▶
Open the hose clamp
⑦
before start-up.
Note
Using a new hose, it may take up to approx. 10 seconds for the coolant to
exit on the handpiece, depending on the feed rate.
4.16.5 Turn on pump and regulate
Requirement
Treatment centre is turned on. The handpiece is connected to the pump via
the pressure line.
▶
Remove the handpiece.
▶
Push down the cross-switch of the foot control for 4 seconds until you hear
the signal.
ð
After activation, "NaCl" cooling can be selected.
Note
The first time the pump is turned on, it takes approximately 10 seconds until
saline solution exits from the handpiece.
The pump does not have any back suction.
▶
Push down the cross switch for 4 seconds until the signal sounds to turn off
the saline pump.
4.16.6 Changing the coolant container
The coolant container can be changed as follows:
▶
Close the hose clamp.